High Performance Dioxin/Furan Immunoassay Kit Insert (IN-DF1)
Last Revised: 12/4/98

Previous Page

I.
Preparation of PCDD/F Sample Extracts for EIA by Solvent Exchange
1.

Locate the vial labeled "0.5 mL neat Triton X-100". Make a stock solution of 100 ppm (0.01% v/v) Triton in methanol by adding 10 µL of Triton X-100 to 100 mL of analytical grade methanol and mixing thoroughly. This stock solution will serve as an EIA negative control and will be used for adding Triton X-100 to extracts prior to solvent exchange and for bringing the Triton X-100 concentration of sample and/or standard EIA tubes to the target level of 10 ppm (final concentration) during EIA analysis.

 

 

The following is a generic extract preparation procedure which is appropriate for samples which have gone through the full cleanup procedure prior to GC/MS analysis (using immunoassay compatible internal standards).

For other samples, consult the appropriate Application Note . Use of this EIA for screening analysis depends upon the same amount of each sample being introduced into the EIA. This quantity (the matrix load in "mg sample equivalent per EIA tube") is dictated by the target sensitivity and must be determined in advance.

Consult Table 3 to determine the appropriate matrix load.

Table 3: TEQ Levels (ppb) at Which Screening Decisions are Made:
(based on three standards and six different matrix loads).

How to use Table 3:
Locate your desired decision level (in ppb TEQ) in the body of the table (the orange cells), then read the "mg Sample Equivalent per EIA tube" at the top of that column. This is the amount of sample you must use per EIA tube. If your decision level occurs in two table columns, use the left one (corresponding to the lower standard in the left column). This will allow less sample to be used, giving less potential for matrix interferences. The comparison of your sample to the appropriate standard in the left column allows you to make the decision you want, as illustrated by the examples below the table.

 
Standards (10 µL/EIA tube)
mg Sample Equivalent per EIA tube
  stock
ppb
pg/tube
0.01
0.1
1 10 100 1000
Std 2
1 10 1000
ppb TEQ
100 10 1 0.1 0.01
Std 3
3.2 32 3200 320 32 3.2 0.32 0.032
Std 4
10 100 10000 1000 100 10 1 0.1
 

Example 1: When using 1 mg sample equivalent per EIA tube, a screening decision at 10 ppb would be made by comparing the sample to Standard 2 (1 ppb 2,3,7,8-TCDD).

Example 2: To make a 10 ppt screening decision by comparison of the sample to Standard 2, a matrix load of 1000 mg sample equivalent per EIA tube would be required.

Example 3: When using 10 mg sample equivalent per EIA tube, a screening decision at 10 ppb would be made by comparing the sample to Standard 4 (10 ppb 2,3,7,8-TCDD).
2.

It is best to exchange only enough sample for immediate analysis, plus the minimum amount to cover waste in pipetting. Leave the rest of the sample in toluene or other less volatile EIA-incompatible solvent.

Redissolving the sample in 30 µL of methanol will allow recovery of one or two 10 µL aliquots for single or duplicate EIA analysis.

3.

Use a clean glass tube or conical vial for evaporating each sample to be tested. Add an appropriate amount of Triton X-100 in methanol solution to each evaporation tube or vial.

If the sample will be redissolved in 30 µL of methanol, then add 30 µL of 100 ppm Triton in methanol.

4. Add standard or sample in volatile solvent such as toluene, hexane, isooctane, or acetone. Solvents with boiling points higher than toluene (111°C) should be avoided if possible.
5.

Evaporate solvent completely under dry gas stream. The original solvent must be completely gone. Recovery will not be adversely affected if the sample remains under the gas stream for a few minutes after reaching dryness.

Complete removal of 100 µL toluene generally takes at least 15 minutes at room temperature. Larger starting volumes may require gentle heat and more time to complete the evaporation. During solvent exchange the Triton X-100 functions as a "keeper", similar to the way dodecane or tetradecane are often used.

When the original solvent is completely evaporated, the PCDD/Fs stay with the Triton, not on the glass, so they are easily redissolved in methanol.

6.

Add methanol to each tube and mix vigorously for 15 to 30 seconds to redissolve the detergent and sample residue. Sonication or longer mixing times have generally not been found to be necessary.

Samples should contain 100 ppm of Triton X-100 when redissolved in methanol. Immediately pipet the redissolved sample into the EIA tube which has been prepared (10 µL per EIA tube as noted in Section J5 below).

This minimizes the evaporative loss of methanol. Add methanol to redissolve the next sample only when you are finished adding the previous sample to the prepared EIA tubes.

If larger reconstitution volumes are used, samples redissolved in methanol can be stored for later analysis if desired.


J.
EIA Analysis of Standards and Prepared Samples
Prepare samples according to the directions above in "Preparation of PCDD/F Sample Extracts by Solvent Exchange" or in the appropriate Application Note. The following steps explain how to analyze your prepared samples using the CAPE Technologies High Performance Dioxin/Furan Immunoassay Kit. For quick reference, a summary of this protocol is provided on a separate sheet (PS-DF1). The number of tubes per run should be limited by the amount of time it takes to add Competitor-HRP Conjugate in step 7 below, and is typically 20 or fewer. This is the largest batch size that can be done on one filling of the 10 mL tip of the Eppendorf Repeater Plus pipettor. Follow precautions in Section E above. Do not expose Substrate to direct sunlight.
1. Place the anti-Dioxin antibody coated tubes in the rack and label them. Put the standard tubes first, from low to high concentration, then the sample tubes.
2. Rinse tubes once by filling each tube with reagent grade or bottled distilled water. Dump water out and tap inverted tubes on absorbent material to remove excess water.
3. Insert the 10 or 12.5 mL pipet tip labeled "water" into the Repeater pipettor and set volume to 500 µL. Dispense one 500 µL aliquot of reagent grade or bottled distilled water into each tube.
4. Using the Repeater pipettor and a 0.5 or 1.0 mL tip (or any other appropriate pipettor) pipet into each EIA tube 50 µL of the stock solution of 100 ppm Triton X-100 in methanol (prepared in section I-1). This solution should be dispensed into the liquid and not onto the side of the tube. Mix the rack of tubes by shaking for 10 seconds. The mixing should be vigorous enough to visibly swirl the liquid around the bottom of the tubes.
5. Using a glass capillary positive displacement pipettor, pipet 10 µL of negative control, standard or sample into each tube. When adding standards and samples to the EIA tubes, the methanol solutions must be dispensed directly into the liquid and not above the liquid surface or onto the side of the tube. Mix the rack of tubes by shaking for 10 seconds after adding the last sample. The mixing should be vigorous enough to visibly swirl the liquid around the bottom of the tubes. Cover the rack of tubes or place in a closed plastic bag or other airtight container with limited headspace. Incubate at room temperature for 12 to 24 hours. The amount of time taken for addition of negative control, standard and sample has little effect on the results because of the long sample incubation. (It is also possible, but not recommended, to use a 2 hour incubation at this point rather than overnight. This will cause a decrease in sensitivity of up to two-fold. Also, results may be affected by proportionally higher variations in incubation time among samples, due to the sample addition process. However, screening decisions would still be made as in Table 3.)
6. Make a wash solution of 100 ppm Triton X-100 in reagent grade or bottled distilled water by adding 10 µL of Triton X-100 to 100 mL of water and mixing thoroughly (this will typically take several minutes on a magnetic stirrer). This amount is sufficient for 20 tubes (20 tubes x 4 washes per tube x 1 mL/wash/tube = 80 mL nominal). Dump or aspirate the EIA tube contents into a suitable waste container. Tap inverted tubes on absorbent material to remove excess liquid. Insert a 50 mL pipet tip into the Repeater pipettor and set volume to 1.0 mL. Dispense one 1 mL aliquot of 100 ppm Triton X-100 in reagent grade or bottled distilled water into each tube. Dump or aspirate the EIA tube contents into a suitable waste container. Repeat this wash step three more times for a total of 4 washes. Be certain to shake or tap out as much wash solution as possible on each wash, especially the last one.
7. Insert the 10 or 12.5 mL pipet tip labeled "conjugate" into the Repeater pipettor and set volume to 500 µL. Dispense one 500 µL aliquot of "Competitor-HRP Conjugate" into each tube. Gently mix all tubes for 10 seconds. Incubate tubes at room temperature for 15 minutes. Timing for this step is the most important of the EIA steps. Rapid and accurate addition of conjugate and consistent incubation times are necessary to maintain equal treatment within and among runs.
8. Repeat the wash procedure described in step 6 above except use reagent grade or bottled distilled water with no detergent added.
9. Insert the 10 or 12.5 mL pipet tip labeled "substrate" into the Repeater pipettor and set volume to 500 µL. Dispense one 500 µL aliquot of "HRP Substrate Solution" into each tube. Mix tubes briefly and incubate at room temperature for 30 minutes.
WARNING: Stop solution is 1N hydrochloric acid. Handle carefully.
10. Insert the 10 or 12.5 mL pipet tip labeled "stop" into the Repeater pipettor and set volume to 500 µL. Dispense one 500 µL aliquot of "Stop Solution" into each tube. The Stop Solution converts the developed color to yellow. If Stop Solution is not added, all tubes will eventually turn dark blue. Read the tubes as soon as possible after stopping; the yellow color is stable for only 30 minutes.
11. To use the Artel Differential Photometer, add at least 1 mL of reagent grade or bottled distilled water to a blank test tube and insert the tube into the left well of the photometer. Dry the outside of each EIA tube, insert tube into the right well of the photometer, and record the absorbance (optical density [OD]) of each tube. Alternatively, read the absorbance of each sample at 450 nm using a tube reader, conventional spectrophotometer, or microplate reader. Consult the Recommended Equipment List (EL-001) for information on how to contact Artel.
../logo.jpg%20(3883%20bytes) Phone - (207) 741-2995
Fax - (207) 741-2996
Email - cape-tech@ceemaine.org
Copyright ©1998 CAPE Technologies L.L.C. All rights reserved.