J.
|
EIA Analysis
of Standards and Prepared Samples |
Prepare samples according to the directions
above in "Preparation of PCDD/F Sample Extracts by
Solvent Exchange" or in the appropriate Application
Note. The following steps explain how to analyze your
prepared samples using the CAPE
Technologies High Performance Dioxin/Furan
Immunoassay Kit. For quick reference, a summary of this
protocol is provided on a separate sheet ( PS-DF1).
The number of tubes per run should be limited by the amount
of time it takes to add Competitor-HRP Conjugate in step
7 below, and is typically 20 or fewer. This is the largest
batch size that can be done on one filling of the 10 mL
tip of the Eppendorf Repeater Plus pipettor. Follow precautions
in Section
E above. Do not expose Substrate
to direct sunlight.
|
1. |
Place the anti-Dioxin antibody coated tubes
in the rack and label them. Put the standard tubes first,
from low to high concentration, then the sample tubes. |
2. |
Rinse tubes once by filling each tube with
reagent grade or bottled distilled water. Dump water out
and tap inverted tubes on absorbent material to remove excess
water. |
3. |
Insert the 10 or 12.5 mL pipet tip labeled
"water" into the Repeater pipettor and set volume
to 500 µL. Dispense one 500 µL aliquot of reagent
grade or bottled distilled water into each tube. |
4. |
Using the Repeater pipettor and a 0.5 or 1.0
mL tip (or any other appropriate pipettor) pipet into each
EIA tube 50 µL of the stock solution of 100 ppm Triton
X-100 in methanol (prepared in section I-1). This solution
should be dispensed into the liquid and not onto the side
of the tube. Mix the rack of tubes by shaking for 10 seconds.
The mixing should be vigorous enough to visibly swirl the
liquid around the bottom of the tubes. |
5. |
Using a glass capillary positive displacement
pipettor, pipet 10 µL of negative control, standard
or sample into each tube. When adding standards and samples
to the EIA tubes, the methanol solutions must be dispensed
directly into the liquid and not above the liquid surface
or onto the side of the tube. Mix the rack of tubes by shaking
for 10 seconds after adding the last sample. The mixing
should be vigorous enough to visibly swirl the liquid around
the bottom of the tubes. Cover the rack of tubes or place
in a closed plastic bag or other airtight container with
limited headspace. Incubate at room temperature for 12 to
24 hours. The amount of time taken for addition of negative
control, standard and sample has little effect on the results
because of the long sample incubation. (It is also possible,
but not recommended, to use a 2 hour incubation at this
point rather than overnight. This will cause a decrease
in sensitivity of up to two-fold. Also, results may be affected
by proportionally higher variations in incubation time among
samples, due to the sample addition process. However, screening
decisions would still be made as in Table
3.) |
6. |
Make a wash solution of 100 ppm Triton X-100
in reagent grade or bottled distilled water by adding 10
µL of Triton X-100 to 100 mL of water and mixing thoroughly
(this will typically take several minutes on a magnetic
stirrer). This amount is sufficient for 20 tubes (20 tubes
x 4 washes per tube x 1 mL/wash/tube = 80 mL nominal). Dump
or aspirate the EIA tube contents into a suitable waste
container. Tap inverted tubes on absorbent material to remove
excess liquid. Insert a 50 mL pipet tip into the Repeater
pipettor and set volume to 1.0 mL. Dispense one 1 mL aliquot
of 100 ppm Triton X-100 in reagent grade or bottled distilled
water into each tube. Dump or aspirate the EIA tube contents
into a suitable waste container. Repeat this wash step three
more times for a total of 4 washes. Be certain to shake
or tap out as much wash solution as possible on each wash,
especially the last one. |
7. |
Insert the 10 or 12.5 mL pipet tip labeled
"conjugate" into the Repeater pipettor and set
volume to 500 µL. Dispense one 500 µL aliquot
of "Competitor-HRP Conjugate" into each tube.
Gently mix all tubes for 10 seconds. Incubate tubes at room
temperature for 15 minutes. Timing for this step is the
most important of the EIA steps. Rapid and accurate addition
of conjugate and consistent incubation times are necessary
to maintain equal treatment within and among runs. |
8. |
Repeat the wash procedure described in step
6 above except use reagent grade or bottled distilled water
with no detergent added. |
9. |
Insert the 10 or 12.5 mL pipet tip labeled
"substrate" into the Repeater pipettor and set
volume to 500 µL. Dispense one 500 µL aliquot
of "HRP Substrate Solution" into each tube. Mix
tubes briefly and incubate at room temperature for 30 minutes. |
WARNING: Stop
solution is 1N hydrochloric acid. Handle carefully.
|
10. |
Insert the 10 or 12.5 mL pipet tip labeled
"stop" into the Repeater pipettor and set volume
to 500 µL. Dispense one 500 µL aliquot of "Stop
Solution" into each tube. The Stop Solution converts
the developed color to yellow. If Stop Solution is not added,
all tubes will eventually turn dark blue. Read the tubes
as soon as possible after stopping; the yellow color is
stable for only 30 minutes. |
11. |
To use the Artel Differential Photometer,
add at least 1 mL of reagent grade or bottled distilled
water to a blank test tube and insert the tube into the
left well of the photometer. Dry the outside of each EIA
tube, insert tube into the right well of the photometer,
and record the absorbance (optical density [OD]) of each
tube. Alternatively, read the absorbance of each sample
at 450 nm using a tube reader, conventional spectrophotometer,
or microplate reader. Consult the Recommended Equipment
List (EL-001)
for information on how to contact Artel.
|
|